Your Avive Connect AED is designed to be stored and function across a range of environments. To stay Emergency Ready, on a long-term basis your device should remain stored between 41°F to 104°F. You can keep track of your device’s temperature status remotely in REALConnect and receive notifications any time it falls outside of its optimal temperature range.
Temperature Specification | Range |
Operating Temperature | 32°F to 122°F (0°C to 50°C) |
Battery Charging Temperature | 32°F to 113°F (0°C to 45°C) |
Short-Term Storage Temperature (up to 14 days) | -4°F to 140°F (-20°C to 60°C) |
Long-Term Storage Temperature | 41°F to 104°F (5°C to 40°C) |
Temperature Warnings
If you see a Temperature Warning for one of your device’s in REALConnect, this means your device has detected that it is outside its operating temperature range of 32°F to 122°F. If your device is stored outside or in a vehicle, it may be more prone to temperature fluctuations, and will send temperature warnings whenever it falls outside of the operating range.
To clear a Temperature Warning, bring your device indoors and store it in a more controlled location, or wait for the environment’s temperature to fall back within 32°F to 122°F. When the device detects that its internal temperature is back within its optimal operating range, it will automatically update in REALConnect. NOTE: The Avive Connect AED has a short-term storage temperature for up to 14 days at -4°F to 140°F so that you’re covered if you are unable to move your device immediately, or if you expect the temperature to soon fluctuate back within the device’s operating range.
Long-term Storage: The long-term storage temperature for your device is 41°F to 104°F. While the device has a short term storage range and is designed to operate in a wider range than this, you should keep it stored between these ranges on a long-term basis.
Extreme Temperatures: If your device exceeds the Short Term storage range of -4°F to 140°F, the AED is too hot or too cold and cannot operate.